Posted Jun 17, 2026
Mito Red Light Inc: Mito Red Light, Super Mobile
Reason stated The near-infrared therapy device contains a lithium-ion battery that may malfunction, causing it to catch fire or burst into flames.
Firm action Beginning on on February 12, 2025, Mito Red Light LLC issued a recall notification to affected consignees in a variety of methods. Mito Red Light is asking consignees to take the following actions: 1. Immediately stop using the mito Mobile Super. 2. Do not charge the device. Disconnect any chargers from the device and rom any power outlet. 3. Store the device away from flammable materials, in a cool dry place,…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 12, 2025
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY,…
FDA recall record Z-2478-2026
Posted May 11, 2026
Technological Medical Advancements LLC: Diowave Laser System
Reason stated Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.
Firm action Starting on 1/9/2026, customer notifications were sent informing customers to install software update V1.81. On 3/11/2026, correction notices were sent to customers informing them to take the following actions: 1) Install Software Update V1.81. Follow the WiFi upgrade instructions to complete the update. The update is free of charge and requires no hardware modifications. 2) Confirm Completion. Once the update is…
- FDA status
- Open, Classified
- Initiated by firm
- Jan 9, 2026
- Quantity in commerce
- 16
- Distribution
- US Nationwide distribution in the states of TX, CA, MI, MD, PA, FL, MN, DC, MI.
FDA recall record Z-2162-2026
Posted Aug 22, 2025
THOR Photomedicine Ltd: THOR Product Name: NovoTHOR
Reason stated The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Firm action On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 24, 2025
- Quantity in commerce
- 80 units (66 US, 14 OUS)
- Distribution
- Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom,…
FDA recall record Z-2416-2025
Posted Aug 22, 2025
THOR Photomedicine Ltd: THOR Product Name: NovoTHOR XL
Reason stated The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Firm action On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 24, 2025
- Quantity in commerce
- 55 units (41 US, 14 OUS)
- Distribution
- Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom,…
FDA recall record Z-2417-2025
Posted Aug 22, 2025
THOR Photomedicine Ltd: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX
Reason stated The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Firm action On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 24, 2025
- Quantity in commerce
- 1 unit
- Distribution
- Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom,…
FDA recall record Z-2418-2025
Posted Aug 22, 2025
THOR Photomedicine Ltd: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX
Reason stated The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Firm action On July 24, 2025, IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions by THOR: A service agent representing THOR will contact you within the next 10 weeks to schedule the fitting of the four high tensile strength steel bolts to the inside of the chassis. The process is likely to take around 3 hours and, of course, there is no fee. Required Actions for NovoTHOR Owners: 1. Please reply to this…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 24, 2025
- Quantity in commerce
- 2 units (1 US, 1 OUS)
- Distribution
- Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom,…
FDA recall record Z-2419-2025