Posted Aug 26, 2026
Olympus Corporation of the Americas: POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action
Reason stated Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
Firm action Olympus notified consignees on 08/18/2026 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were notified of the issue, hazard involved, and that the affected device could still be used. Consignees were instructed to review the recall notification and current inventory for the affected lot number, discard the outer box for any affected units on hand, and attach the recall notification letter to the inner…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 18, 2026
- Quantity in commerce
- 335 units
- Distribution
- Worldwide - US Nationwide distribution in the states of Kansas and Florida. The countries of Australia, European Union, Japan.
FDA recall record Z-3035-2026
Posted Aug 3, 2026
Stryker Corporation: STRYKER PKG., SINGLE SPIKE TUBE SET, Catalog Number 0250070530
Reason stated There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
Firm action Stryker issued an URGENT: Medical Device Recall notice to its consignees on 7/6/2026 via USPS Certified Mail. The notice explained the issue, potential risks, and requested the following: "Actions needed: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product using Attachment B. 3. If affected product is found a.…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 8, 2026
- Quantity in commerce
- 11,857 Units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Brazil, Canada, Chile, France, Germany, Indonesia, Italy, Mexico, Netherlands, New Zealand, Philippines, South Korea, Spain,…
FDA recall record Z-2865-2026
Posted Aug 3, 2026
Stryker Corporation: STRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540
Reason stated There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
Firm action Stryker issued an URGENT: Medical Device Recall notice to its consignees on 7/6/2026 via USPS Certified Mail. The notice explained the issue, potential risks, and requested the following: "Actions needed: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product using Attachment B. 3. If affected product is found a.…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 8, 2026
- Quantity in commerce
- 589 Units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Brazil, Canada, Chile, France, Germany, Indonesia, Italy, Mexico, Netherlands, New Zealand, Philippines, South Korea, Spain,…
FDA recall record Z-2866-2026
Posted Jul 15, 2026
Maquet Cardiovascular, LLC: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000
Reason stated Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
Firm action Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 2, 2026
- Quantity in commerce
- 205 units
- Distribution
- US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
FDA recall record Z-2741-2026
Posted Apr 17, 2026
Olympus Corporation of the Americas: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm
Reason stated Potential for detachment of a distal tip component of the device during use.
Firm action Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 23, 2026
- Quantity in commerce
- 830 units
- Distribution
- International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
FDA recall record Z-1851-2026
Posted Apr 17, 2026
Olympus Corporation of the Americas: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm
Reason stated Potential for detachment of a distal tip component of the device during use.
Firm action Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 23, 2026
- Quantity in commerce
- 365 units
- Distribution
- International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
FDA recall record Z-1852-2026
Posted Apr 17, 2026
Olympus Corporation of the Americas: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm
Reason stated Potential for detachment of a distal tip component of the device during use.
Firm action Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 23, 2026
- Quantity in commerce
- 3,360 units
- Distribution
- International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
FDA recall record Z-1853-2026
Posted Apr 17, 2026
Olympus Corporation of the Americas: Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm
Reason stated Potential for detachment of a distal tip component of the device during use.
Firm action Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 23, 2026
- Quantity in commerce
- 490 units
- Distribution
- International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
FDA recall record Z-1854-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 106 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1638-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 164 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1639-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 444 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1640-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 2648 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1641-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 93 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1642-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 7 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1643-2026
Posted Mar 25, 2026
Olympus Corporation of the Americas: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps
Reason stated Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Firm action On February 25, 2026 URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: 1. Examine inventory and quarantine any affected devices 2. Cease usage of the product with immediate effect 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization 4. Even if you no longer have the device, Olympus requests that you acknowledge receipt of this…
- FDA status
- Open, Classified
- Initiated by firm
- Feb 25, 2026
- Quantity in commerce
- 326 units
- Distribution
- Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
FDA recall record Z-1644-2026
Posted Nov 26, 2025
Stryker Corporation: NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP
Reason stated Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Firm action On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.
- FDA status
- Open, Classified
- Initiated by firm
- Oct 31, 2025
- Quantity in commerce
- 39,148
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
FDA recall record Z-0599-2026
Posted Nov 26, 2025
Stryker Corporation: NICO Myriad Illumination Pack Product No
Reason stated Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Firm action On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.
- FDA status
- Open, Classified
- Initiated by firm
- Oct 31, 2025
- Quantity in commerce
- 39,148
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
FDA recall record Z-0600-2026
Posted Nov 26, 2025
Stryker Corporation: NICO BrainPath
Reason stated Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.
Firm action On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.
- FDA status
- Open, Classified
- Initiated by firm
- Oct 31, 2025
- Quantity in commerce
- 39,148
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
FDA recall record Z-0601-2026
Posted Oct 14, 2025
Reason stated Manufacturing defects in devices manufactured from 4/1/25 to 8/20/25 can on rare occasions result in adverse events for patients.
Firm action An URGENT-Field Safety Corrective Action - XERF Tips notification dated 9/4/25 was mailed to consignees notifying them of this recall. Consignees are instructed to immediately stop use of affected devices and to return all units utilizing the provided shipping label. Consignees with any questions can contact Customer Service at 1-800-886-2966 or by email at [email protected].
- FDA status
- Open, Classified
- Initiated by firm
- Sep 4, 2025
- Quantity in commerce
- 7,490 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, IL, MA, NJ, NY, PA, TX, and UT. The countries of Canada, South Korea, Hong Kong, Japan, Singapore, Vietnam, Myanmar, Philippines, Taiwan,…
FDA recall record Z-0146-2026
Posted Sep 26, 2025
Bard Peripheral Vascular Inc: Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation…
Reason stated Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification…
Firm action On 8/21/2025, correction notices were mailed or emailed to Health Care Professionals (HCPs), Risk Manager, Head Nurse, Nurse Manager, Medical Device Safety Officer, Procurement Officer, Office Manager, and Distributors, were advised of the following: If the generator is operating with affected software and is connected to a catheter and the software check fails, a red "X" will be displayed on the generator screen…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 21, 2025
- Quantity in commerce
- 1,725
- Distribution
- Worldwide - US Nationwide distribution including in the states of NJ, NY, FL, LA, VA, CA, TX, NV, PA, NE, TN, IL, AZ, SC, AK, GA, CO, HI, OH, NC, WI, WV, MD, MO, AL, OK, MS, PR, KS, UT, CT, IA, MI,…
FDA recall record Z-2638-2025
Posted Sep 25, 2025
Maquet Cardiovascular, LLC: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No
Reason stated Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
Firm action Getinge Field Representatives contacted impacted accounts by telephone on 8/6/25 to notify them of this recall action and provide instructions for quarantine and return of affected devices. Consignees were asked to stop all visual inspection and return all HemoPro 3 units. Returns can be coordinated by calling or emailing Michele Kanis for an RMA at [email protected] or (408)635-0647. Consignees are to…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 6, 2025
- Quantity in commerce
- 98 units
- Distribution
- US Nationwide distribution in the states of FL, IL, MD & TX.
FDA recall record Z-2636-2025
Posted Aug 19, 2025
Olympus Corporation of the Americas: Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP)
Reason stated Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Firm action Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 25, 2025
- Quantity in commerce
- 155 units
- Distribution
- US Nationwide distribution.
FDA recall record Z-2338-2025
Posted Aug 19, 2025
Olympus Corporation of the Americas: Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410
Reason stated Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
Firm action Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further…
- FDA status
- Open, Classified
- Initiated by firm
- Jul 25, 2025
- Quantity in commerce
- 434 units
- Distribution
- US Nationwide distribution.
FDA recall record Z-2339-2025
Posted Apr 18, 2025
Maquet Cardiovascular, LLC: The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope
Reason stated There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from…
Firm action An URGENT MEDICAL DEVICE NOTIFICATION dated March 2025 was mailed to consignees. The notification provides instructions to users on how to proceed with a procedure if a ceramic C-Ring break occurs. If a consignee has a C-Ring break they are asked to save the device and complete the provided return form so Maquet Cardiovascular can conduct an investigation. Maquet instructs consignees to forward the notification to…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 20, 2025
- Quantity in commerce
- 22,244 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.
FDA recall record Z-1622-2025
Posted Apr 16, 2025
Maquet Cardiovascular, LLC: Maquet VH-3010 Power Supply
Reason stated Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Firm action An URGENT MEDICAL DEVICE REMOVAL notification dated March 2025 was mailed to consignees. The notification instructs consignees to examine their inventory for impacted serial numbers. For power supplies whose serial numbers are not listed in the provided Appendix A of recalled devices, devices can continue to be used. If power supplies with affected serial numbers are identified as being in stock, the associated…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 10, 2025
- Quantity in commerce
- 110 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan,…
FDA recall record Z-1577-2025
Posted Apr 9, 2025
MEDLINE INDUSTRIES, LP - Northfield: MEDLINE ReNewal Coblator II Evac 70 Xtra Plasma Wand,w/Integrated Saline Line & Integrated Cable, Item Number
Reason stated Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Firm action Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 5, 2025
- Quantity in commerce
- 738 units
- Distribution
- US Nationwide distribution
FDA recall record Z-1546-2025
Posted Apr 9, 2025
MEDLINE INDUSTRIES, LP - Northfield: MEDLINE ReNewal Coblator II ENT Evac 70 Xtra HP,w/Integrated Cable (Blue/Grey), Number
Reason stated Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Firm action Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 5, 2025
- Quantity in commerce
- 020 units
- Distribution
- US Nationwide distribution
FDA recall record Z-1547-2025
Posted Apr 9, 2025
MEDLINE INDUSTRIES, LP - Northfield: MEDLINE ReNewal ENT Coblator II PROcise XP Wand,w/Integrated Cable Suction & Saline (Blue), Item Number
Reason stated Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Firm action Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 5, 2025
- Quantity in commerce
- 315 units
- Distribution
- US Nationwide distribution
FDA recall record Z-1548-2025
Posted Apr 9, 2025
MEDLINE INDUSTRIES, LP - Northfield: MEDLINE ReNewal ENT Coblator II PROcise mAx,W/ Integrated Cable Suction & Saline (Blue), Item Number
Reason stated Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Firm action Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response…
- FDA status
- Open, Classified
- Initiated by firm
- Mar 5, 2025
- Quantity in commerce
- 224 units
- Distribution
- US Nationwide distribution
FDA recall record Z-1549-2025
Posted Feb 11, 2025
Stryker Corporation: stryker Precision Thin, Catalog Number
Reason stated Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Firm action Stryker issued an URGENT MEDICAL DEVICE recall notice to its consignees on 01/08/2025 via FedEx. The notice explained the issue, potential risk, and requested the affected items be quarantined pending return. Stryker will issue a shipping label to the consignee once the response form has been received. Distributors were directed to forward the notice to all those to whom the affected devices were distributed.
- FDA status
- Open, Classified
- Initiated by firm
- Jan 8, 2025
- Quantity in commerce
- 622 units
- Distribution
- Worldwide distribution.
FDA recall record Z-1138-2025
Posted Feb 11, 2025
Stryker Corporation: stryker Precision Thin, Catalog Number
Reason stated Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Firm action Stryker issued an URGENT MEDICAL DEVICE recall notice to its consignees on 01/08/2025 via FedEx. The notice explained the issue, potential risk, and requested the affected items be quarantined pending return. Stryker will issue a shipping label to the consignee once the response form has been received. Distributors were directed to forward the notice to all those to whom the affected devices were distributed.
- FDA status
- Open, Classified
- Initiated by firm
- Jan 8, 2025
- Quantity in commerce
- 238 units
- Distribution
- Worldwide distribution.
FDA recall record Z-1139-2025
Posted Jan 6, 2025
Stryker Corporation: 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V…
Reason stated Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.
Firm action On December 3, 2024, the firm notified affected consignees via "Urgent Medical Device Field Action" letters dated December 2, 2024. Customers were instructed to locate Neptune S Waste Management Systems and to quarantine affected products at their facility; to contact their Stryker Sales Rep to schedule the replacement of the affected device; to complete Business Reply Form after scheduling replacement and email…
- FDA status
- Open, Classified
- Initiated by firm
- Dec 3, 2024
- Quantity in commerce
- 14 (Expanded to 21 units on 10/9/25)
- Distribution
- US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.
FDA recall record Z-0807-2025
Posted Jan 3, 2025
Maquet Cardiovascular, LLC: VasoView HemoPro 2 Endoscopic Vessel Harvesting System
Reason stated Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
Firm action Getinge notified consignees on 12/09/2024 via Urgent Medical Device Correction letter posted digitally on Getinge.com and sent to consignees via letter. Consignees were not instructed to return affected units, but instead to review the IFU information provided with the letter to minimize over-delivery of energy and mitigate risks. Getinge instructed consignees to notify customers if further distributed, as well as…
- FDA status
- Open, Classified
- Initiated by firm
- Dec 9, 2024
- Quantity in commerce
- 418,076 units (326,608 US, 91,468 OUS)
- Distribution
- Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran,…
FDA recall record Z-0711-2025
Posted Jan 3, 2025
Maquet Cardiovascular, LLC: Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System
Reason stated Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
Firm action Getinge notified consignees on 12/09/2024 via Urgent Medical Device Correction letter posted digitally on Getinge.com and sent to consignees via letter. Consignees were not instructed to return affected units, but instead to review the IFU information provided with the letter to minimize over-delivery of energy and mitigate risks. Getinge instructed consignees to notify customers if further distributed, as well as…
- FDA status
- Open, Classified
- Initiated by firm
- Dec 9, 2024
- Quantity in commerce
- 11,101 units US
- Distribution
- Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran,…
FDA recall record Z-0712-2025