Posted Oct 16, 2026 · Ultrasound
Reason stated Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
Firm action On 9/24/2026, correction notices were mailed to customers who were asked to do the following: Until software update is available and installed: " Do not use the affected systems outside of areas with readily accessible Wi-Fi/Cellular (e.g. prehospital transport (ambulance, helicopter) or use outside of a traditional healthcare setting (care in the field). " Always have backup ultrasound devices and/or modalities…
- FDA status
- Open, Classified
- Initiated by firm
- Sep 24, 2026
- Quantity in commerce
- 13,206
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV,…
FDA recall record Z-0152-2027
Posted Oct 9, 2026 · TMS
Neuronetics, Inc.: NeuroStar Advanced Therapy System
Reason stated Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.
Firm action NeuroStar began notification to consignees on about 08/24/2026 via email. Consignees were instructed to check their NeuroStar System for Console and Coil for affected serial numbers and complete and return the response form. A NeuroStar Field Service Engineer will contact you to coordinate product inspection at no expense to you. If the HP Coild Assembly is not affected, no action is required. If it is determined…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 24, 2026
- Quantity in commerce
- 179 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Japan and South Korea.
FDA recall record Z-0142-2027
Posted Sep 25, 2026 · Ultrasound
Volcano Corp: Philips IntraSight Plus
Reason stated Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.
Firm action On August 6, 2026, Philips issued a Urgent Medical Device Correction to affected consignees. Philips ask consignees to take the following actions: 1. Contact your local Philips Service representative in advance of any changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment, as such changes may affect the Study ID field. Your Philips Service representative will help…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 6, 2026
- Quantity in commerce
- 19 systems
- Distribution
- Worldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA and the countries of Belgium, Denmark, Germany, Netherlands, Poland, Spain, Sweden, United Kingdom.
FDA recall record Z-3276-2026
Posted Sep 18, 2026 · Ultrasound
Philips Ultrasound, LLC: Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
Reason stated Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Firm action On August 7, 2026, Philips issued a Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions: 1. Due to possibility of a weak PW Doppler Signal on EPIQ Elite running 14.0 SW, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1. Please note that this issue…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 7, 2026
- Quantity in commerce
- 219 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany,…
FDA recall record Z-3249-2026
Posted Sep 4, 2026 · Lasers
Reason stated Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Firm action Firm began notifying customers on 8/6/26. Customers were instructed to discontinue use of previous parameters and utilize the new revised clinical cryogen parameters. Updated cryogen parameters were provided with the letter. If you have questions or need assistance, please contact Clinical Support at 844-365-5060 or via email at [email protected].
- FDA status
- Open, Classified
- Initiated by firm
- Aug 6, 2026
- Quantity in commerce
- 1065
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada and European Union.
FDA recall record Z-3110-2026
Posted Aug 26, 2026 · RF and Skin Tightening
Olympus Corporation of the Americas: POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action
Reason stated Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
Firm action Olympus notified consignees on 08/18/2026 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were notified of the issue, hazard involved, and that the affected device could still be used. Consignees were instructed to review the recall notification and current inventory for the affected lot number, discard the outer box for any affected units on hand, and attach the recall notification letter to the inner…
- FDA status
- Open, Classified
- Initiated by firm
- Aug 18, 2026
- Quantity in commerce
- 335 units
- Distribution
- Worldwide - US Nationwide distribution in the states of Kansas and Florida. The countries of Australia, European Union, Japan.
FDA recall record Z-3035-2026
Posted Aug 3, 2026 · RF and Skin Tightening
Stryker Corporation: STRYKER PKG., SINGLE SPIKE TUBE SET, Catalog Number 0250070530
Reason stated There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
Firm action Stryker issued an URGENT: Medical Device Recall notice to its consignees on 7/6/2026 via USPS Certified Mail. The notice explained the issue, potential risks, and requested the following: "Actions needed: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product using Attachment B. 3. If affected product is found a.…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 8, 2026
- Quantity in commerce
- 11,857 Units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Brazil, Canada, Chile, France, Germany, Indonesia, Italy, Mexico, Netherlands, New Zealand, Philippines, South Korea, Spain,…
FDA recall record Z-2865-2026
Posted Aug 3, 2026 · RF and Skin Tightening
Stryker Corporation: STRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540
Reason stated There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.
Firm action Stryker issued an URGENT: Medical Device Recall notice to its consignees on 7/6/2026 via USPS Certified Mail. The notice explained the issue, potential risks, and requested the following: "Actions needed: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product using Attachment B. 3. If affected product is found a.…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 8, 2026
- Quantity in commerce
- 589 Units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Brazil, Canada, Chile, France, Germany, Indonesia, Italy, Mexico, Netherlands, New Zealand, Philippines, South Korea, Spain,…
FDA recall record Z-2866-2026
Posted Jul 15, 2026 · RF and Skin Tightening
Maquet Cardiovascular, LLC: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000
Reason stated Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
Firm action Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative…
- FDA status
- Open, Classified
- Initiated by firm
- Jun 2, 2026
- Quantity in commerce
- 205 units
- Distribution
- US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.
FDA recall record Z-2741-2026
Posted Jun 24, 2026 · Lasers
IPG Medical Corporation: Thuvera Laser Console
Reason stated Software anomaly that causes a false display of error code 5018.
Firm action An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their…
- FDA status
- Open, Classified
- Initiated by firm
- May 5, 2026
- Quantity in commerce
- 198 units
- Distribution
- US Nationwide distribution in the state of MA.
FDA recall record Z-2554-2026